2004B11 Explain how a pulse oximeter works. Do NOT discuss the potential sources of error.

 

List:

·     Physics

·     Device

 

Physics:

Beer-Lambert principle

·     Absorption ∝ concentration of the absorbing substance in the medium (Beer’s law)

·     Absorption ∝ distance travelled through the medium (Lambert’s law)

·     Describes the basis for, but not the functioning of, the pulse oximeter

Known absorption spectra

·     940nm: O2Hb > HHb

·     660nm: O2Hb < HHb

 

 

Device:

Components

·     Probe on the finger, nose, ear

·     Two monochromatic LEDs on one side, rapid on and off

·     Photodetector on the other side

·     Electrical cable

·     Processor

·     Display: SpO2 and plethysmograph

Function

(1)Pulsatile signal isolated

·     Excludes venous and tissue absorption

(2)Ratio of absorbances calculated

·     Ratio = (pulsatile660 / non-pulsatile660) / (pulsatile940 / non-pulsatile940)

(3)Ratio correlated with SaO2 derived from experiments

·     R 0.4 -> 100%

·     R 1 -> 85%

·     R 2 -> 50%

 

 